- Applicable Conditioner Cards
-
CDV-40B (-F), CVM-41A, DPM-42B (-F), DPM-42B-I (-F), CTA-40A, CFV-40A, CCA-40A (-F), CDV-44AS, CDA-44AS, CDA-45AS, AD-40AS (-F)
- Analog Input
-
See spec of respective conditioner cards.
- Digital I/O
-
I/O points: Max .8
I/O Settings: Sets every bit to digital input, digital output, or remote-control input with the same common.
*Remote-control input: Measuring start, stop, balance conducting, etc.
Input modes: Isolated, TTL level input
Input voltage: Max. 5 VDC
Isolation modes: Digital isolation
Output modes:Isolated open-collector output (With 10 kΩ built-in pull-up resistors)
Output voltage: 5 VDC
Output current: Max. 25 mA (per point)
- Voice Memo Input
-
1 channel (To be recorded together with measuring data.) An optional remote control unit RCU-42 is required.
Sampling Frequencies
- Sampling Mode
-
1-2-5 series
1 Hz to 200 kHz for up to 32-channel data acquisition
1 Hz to 100 kHz for up to 64-channel data acquisition
1 Hz to 50 kHz for up to 80-channel data acquisition
1 Hz to 10 kHz for real-time synchronous data processing
2n series
2 to 131.072 Hz for up to 32-channel data acquisition
2 to 65.5536 Hz for up to 64-channel data acquisition
2 to 32.768 Hz for up to 80-channel data acquisition
2 to 8.192 Hz for real-time synchronous data processing
*The condition is using a 16-bit A/D converter.
*As for a 24-bit A/D converter, the channels above will be half of each channel.
- Display
-
I
- Operation Keys
-
Keys (REC, STOP, BAL, and OPT) are operated by touchscreen, keys on EDX, external keyboard, or mouse
- Output Control Connectors
-
CONT.IN, CONT.OUT
(For remote control unit, or synchronous operation)
- External I/O Connectors
-
External triggers: TRG IN, TRG OUT
External clock: CLK IN, CLK OUT
(Output at any frequency division ratio)
- External Interfaces
-
USB type keyboard, USB type mouse
VGA connector for monitor
USB I/F: 3 USB 2.0 in front, 2 USB 3.0 in rear
LAN I/F: 100/1000BASE-T
- Power Supply
-
100 to 240 VAC and 10 to 30 DC, dual supply
Built-in batteries against momentary power failure
- Operating Temperature
-
0 –
40
°C
- Operating Humidity
-
20 –
80
% RHI
- Storage Temperature
-
-20 –
60
°C
- Vibration Resistance
-
EDX-5000A-64-HE, EDX-5000A-64-SE
49.0 m/s2 (5 G), 5 to 55 Hz (When not operating)
29.4 m/s2 (3 G), 5 to 55 Hz (When operating)
EDX-5000A-64-HE:
9.8 m/s2 (1 G), 10 to 200 Hz (When operating)
EDX-5000A-64-SE:
19.6 m/s2 (2 G), 10 to 200 Hz (When operating)
- Impact Resistant
-
196.1
m/s2I
- Optional Functions
-
CAN data acquisition (Input channels: Max 512)
GPS data acquisition (Tracing data acquisition, time synchronous function)
- Online Control
-
Controlled by DCS-100A
- Low Voltage Directive
-
EN61010-1
I
- RoHS-Directive
-
EN50581
- Standard Accessories
-
AC power cable P-18 (With 2-pin conversion plug
CM-39), DC power cable P-70, gound wire P-72, simplified instruction manual, CD-R (Instruction manual), EDX accessory bag, AC fuse, DC fuse, soldering plug for DIO, shell case for DIO plug
- Optional accessories
-
Synchronous cable N-128, remote control unit RCU-42A , belt hook, dummy panel for EDX EDX1P-DUMMY, EDX-2000-DUMMY
Software Specifications
Measuring Condition Settings
- Measuring Channel Conditions
-
Measurement ON/OFF, measuring modes, range, HPF, LPF, digital HPF, digital LPF, balance ON/OFF, CAL, calibration coefficient, offset, offset zero ON/OFF, unit, channel name, measuring range, rated capacity, rated output, upper-limit check value, lower-limit check value, numeric display digits (Selects any items to display), inner sensitivity register ON/OFF, input cable, cable compensation value
- Recordable Data
-
Up to remaining disk space of built-in memory
- Manual Measurement
-
Recording from REC to STOP, or preset data starts from REC
- Interval Measurement
-
Automatic recording according to designated starting time and recording interval
- Trigger Measurement
-
• Common trigger conditions
End trigger: Settable
Pre-trigger: Max. 2097152 points per channel
Post-trigger: Max. 2097152 points per channel
(When A/D is set to 24 bits)
Trigger data correlates with the number of channels.
• Analog trigger conditions
Trigger channels: Any 1 analog channel
Trigger level: Sets in physical quantity.
Trigger slope: Up, down
• Digital trigger conditions
Trigger bit: Any 1 bit
Trigger level: 0, 1
• External trigger conditions
Trigger slope: Up, down
• Complex trigger conditions
Trigger source: Selection of any 4 analog/
digital channels, an external
trigger channel, or a manual
trigger channel
AND/OR: Be used for analog
trigger, digital trigger and
external trigger.
Trigger level: An physical quantity is set for
the analog channel, and
0 or 1 for the digital channel.
Trigger slope: Up, down
- TEDS
-
Reads sensor's information and sets to channel condition automatically.
Measuring Operations
- Real-time processing
-
Simultaneous monitoring and recording of data
The sampling frequencies up to 10 kHz are available.
Video Data Acquisition with a Web Camera
- Camera
-
DirectX compatible Web camera
(Recognized by the OS as an imaging device)
- Number of Cameras
-
1
- Resolution
-
Max. 640 x 480
- Frame rate
-
Max. 30 fps
- Saving File Format
-
AVI format
Resolution and frame rate depend on the camera.
The Web camera is optional.
- Measuring Conditions during Recording Video
-
Manual mode, manual mode (Data points preset)
- Arithmetic Operation
-
Calculation channels: Max. 64
Calculation conditions: ON/OFF, operators (Max. 200 characters), unit, numeric display digits, channel name (Max. 40 characters), calculation zero, and calculation zero value
Applicable Operators and Constants
+, -, ×, /, ^[power], P[I π], ()
Functions:
SQR Square root
ABS Absolute value
SIN Sine
COS Cosine
TAN Tangent
ASIN Arc sine (Return value: Radian)
ACOS Arc cosine (Return value: Radian)
ATAN Arc tangent (Return value: Radian)
DSIN (Return value: Angle)
DCOS (Return value: Angle)
DTAN (Return value: Angle)
LOG Common logarithm
LN Natural logarithm
EXP Exponent
HMX Max. principal strain
HMN Min. principal strain
HSM Max. shearing strain
SMX Max. principal stress
SMN Min. principal stress
SSM Max. shearing stress
DEG Principal strain direction
FFT Analysis
- Analysis Types
-
Linear spectrum, power spectrum, cross spectrum, auto-correlation, and cross-correlation
- Window Functions
-
OFF, Hamming, Hanning, Fejer, Blackman, and Gaussian
- Analyzed Data Points
-
256, 512, 1024, 2048, 4096, and 8192
- Analysis Channels
-
4 channels/window
- Saving File Format
-
Kyowa standard file format (KS2)
KS2 file version: 01.06
Monitor
- Y-time Graph
-
X axis indicates the time, and Y axis the physical values of measurement for a maximum of 8 channels.
- X-Y Graph
-
X and Y axes indicate the measured data, max. 4 channels
- Bar Graph
-
Up to 80 channels are contained in a graph.
Displays max. and min. values are possible.
- Circle Meter
-
Any one channel is displayed in a circle meter.
- Numeric Display
-
Up to 80 channels are listed.
Displays max. and min. values of each channel are possible.
- FFT Graph
-
Up to 4 channels of analyzed results are displayed.
- Web Camera
-
Displays captured images.
- Over-input
-
Displays the over-input values, background, etc. in red.
- Graph Scale
-
Capable of displaying auto-scale value on the Y-time graph (Y axis), X-Y graph (X, Y axes), and bar graph (Y axis).
- Cursors
-
Available on Y-time graph, X-Y graph, physical values appear at the cursor positions
- Windows
-
Up to 12 different graph windows are available at the same time.
- Saving Images
-
Saves each graph as an image file.
- Synchronous Operation
-
Standalone, sync host, or sync guest is selected
- Others
-
Inner, external oscillators switching, operating beep, balance reference value, A/D bits switching, and optional switches selection
GPS/Multichanel CAN optional Functions
- Can Data Recording
-
Up to 512 channels of CAN data can be recorded
for each unit.
(CAN data is saved in E4A file format.)
*Use the ELA-55A for CAN condition settings.
- Can Trigger
-
Start/stop recording data by using analog triggers,
ditital triggers, and complex triggers.
*Use the ELA-55A for CAN condition settings.
- Graph Display
-
Display numerical and analog data in graphs simultaneously.
Environmental characteristics
- Point ZERO Measurement
-
Able to start manual measurement at zero second (at the moment when msec is zero) using the time data received from GPS satellites.
- Interval Measurement
-
Able to start interval measurement at zero second (at the moment when msec is zero) using the time data received from GPS satellites.
- GPS Data Recording
-
Able to monitor and record GPS data (latitude, longitude, speed, bearing, etc.).
GPS data is saved in NMEA file format.