High technologies cultivated in weight control of large scale airplanes and original ideas are incorporated into the revolutionary thin design of LCTA-A series load cells. Besides accuracy the integrated flat design and rubber attachment enables use with the top and bottom fixed and provided excellent buffer function and ease of use.
Features:
- Rated Capacity: 500 N to 3 kN
- Advanced thin design - When compared with conventional load cells, the height is 1/2 to 1/3. Space saving expands the application range.
- Usable with top and bottom fixed - Optional dedicated rubber attachment enables fixing the top and bottom with bolts, thereby making it possible to design the system with no tension rod or stay rod used.
- Strong against lateral load - Safety factor is 3 to 5 times higher than conventional load cells. Endures lateral loads up to 20% of the rated capacity.
- Excellent impact/vibration resistance - Rubber attachment attenuates impact energy and lessens the effects of thermal expansion of system and moment of fixed section.
- Easy installation - Rubber attachment facilitates installation with less care about parallelism.
- Wide variation in accuracy and reliability - Varieties of accuracies and outputs are available, enabling configuration of the most suitable system for each individual application.
- Compatibility with peripheral equipment - Since wirings are the same as conventional load cells, peripheral instruments such as amplifiers can be easily connected.
TEDS installed versions can also be manufactured. Inquiries are welcome.
Specifications
Performance
- Nonlinearity
-
+/- 0.05
% RO
- Hysteresis
-
+/- 0.05
% RO
- Repeatability
-
0.03
% RO
- Rated Output
-
2 +/- 0.2%
mV/V
Environmental Characteristics
- Safe Temperature Range
-
-20 –
70
°C
- Compensated Temperature Range
-
-10 –
60
°C
- Temperature Effect on Zero
-
+/- 0.01
% RO/ºC
- Temperature Effect on Output
-
+/- 0.01
%/°C
Electrical Characteristics
- Safe Excitation Voltage
-
20 VDC
VDC
- Recommended Excitation Voltage
-
1 –
10
VI
- Input Resistance
-
350 Ω +/- 1.5%
- Output Resistance
-
350 Ω +/- 1.5%
- Cable
-
Cable
Wires: 4 (0.5 mm2)
Isolation: Chloroprene
Diameter: 8.5 mm
Length: 5 m
Bared at the tip
*Shield wire is not connected to the case.
Mechanical Properties
- Safe Overloads
-
150
%
- Weight
-
1.1
kg
- ultimate-lateral-load-per
-
20
%I
- Weight
-
1.1
kg
- material
-
Aluminum alloy
Your search returned no results.
D
PRC03-12A10-7M
P12-7 Tajimi Plug
Request
S
PRC03-12A10-7M
P12-7 Tajimi Plug
- Contact Count
-
7
- material
-
Housing: Zinc Alloy
Contact: Copper
Contact Plating: Gold
- Operating Temperature
-
-25 –
85
°C
S
D
RA02-100K
Rubber attachment for LCTA-A-500N, LCTA-A-800N, LCTA-A-1KN.
Request
S
RA02-100K
Rubber attachment for LCTA-A-500N, LCTA-A-800N, LCTA-A-1KN.
- Models
-
LCTA-A-500N, LCTA-A-800N, LCTA-A-1KN
S
D
RA02-300K
Rubber Attachment for LCTA-A-2KN, LCTA-A-3KN.
Request
S
RA02-300K
Rubber Attachment for LCTA-A-2KN, LCTA-A-3KN.
- Models
-
LCTA-A-2KN, LCTA-A-3KN.
S
D
COV03-300K
Stainless Steel Cover for LCTA-A
Request
S
COV03-300K
Stainless Steel Cover for LCTA-A
S
D
JB4LC-US
Junction Box for 4 load cells
Request
S
JB4LC-US
Junction Box for 4 load cells
S