Messziele
Strain (Gage, transducer), voltage, thermocouples, pulse (F/V), piezoelectric acceleration (Built-in amplifier), CAN signals
Sprachmemoeingabe
1 channel (Input voice memo data is recorded together with the measurement data.)
Use remote control unit RCU-42A. (Optional accessory)
Use the Data Analysis Software DAS-200A (optional accessory) to play back recorded voice memos.
Sampling
All channels synchronously
Sampling Mode : Normal: All channels collected using the same sampling frequency.
Dual: High-speed or low-speed, collected using 2 types of sampling frequency set for each channel
Sampling Frequencies : Normal Mode
1-2-5 series 1 Hz to 100 kHz
1 Hz to 2 kHz When using CAN-41A
2n series: 2 Hz to 65536 Hz
2 Hz to 2048 Hz When using CAN-41A
Dual Mode High-speed sampling (Expressed as Sf)
1-2-5 series 1 Hz to 100 k Hz
2n series: 2 Hz to 65536 Hz
Low-speed sampling (Expressed as "Ss")
1-2-5 series The division frequencies from
high-speed sampling, and Ss ≤ Sf/4
2n series: The division frequencies from
high-speed sampling, and Ss ≤ Sf/4
Akquisekanal
Normalmodus
Max. 32 channels, 320 k/I (I is the integer part of the set sampling frequency.)
Dualbetrieb
Max. 64 channels, 320 k/I (I is the integer part of the set sampling frequency.)
Digitalfilter
Butterworth filter (IIR)
Type of filter : LPF, HPF
Order of a Filter : 1 to 8
Amplitude ratio at cutoff point : -3 dB
Attenuation : -6 x N dB/oct. (N is filtering times )
Simultaneously use with built-in LPF possible.
Application on CAN data not possible.
Datenaufzeichnung
CF cards, 128 MB to 16 GB (Kyowa recommended items)
Maximum data file size (Available for data acquisition):
4 GB for 1-time measurement
1 GB for repetitive measurement, 2 times or more
Bedienungsschalter
REC/PAUSE: Start/pause data recording.
STOP: Stop data recording.
BAL.: Implement balance (Balance adjustment)
LOAD: Read and configure conditions from CF card
OPT. : Execute arbitrary configured functions
ID: EDX identifier configuration
POWER: Power switch
USB/LAN: Communications I/F, switchable
Externer Steuerungsanschluss
CONT. IN
(Remote control, for synchronous operation)
Schnittstelle
I
Umgebungsbedingungen
Online : From the PC via LAN or USB port
Offline : By reading from the CF card which has measuring conditions.
Written with the data acquisition software DCS-100A
Speicher Bedingungen
Recording of conditioner configuration conditions and measurement conditions within the EDX built- in nonvolatile memory.
Immediate start of data collection using the previously configured measurement conditions after power-on is possible.
Messmodus
Manual measurement/trigger measurement/interval measurement
Manual Measurement : Data recording is manually started or stopped when data is recorded to a preset number of measured data.
Manual mode allows recording of voice memo during data recording.
Trigger Measurement : Data recording is automatically started when the preset trigger condition is satisfied.
*No CAN data of CAN-41A is used as the trigger condition.
Interval Measurement : Automatically recording functions based upon previously-set interval conditions
Combination with measured mode when in dual sampling
Combination with Measured Mode when in Dual Sampling : (High-speed sampling channel - Low-speed sampling channel)
Manual setting - Manual setting
Trigger setting - Manual setting
Trigger setting - Interval setting
Interval setting - Interval setting
Manueller Start/Stop der Datenaufzeichnung
PC, operation switch (panel screen),
or using dedicated remote control
Gleichgewichtseinstellung
Strain input channel balance adjustment is performed using PC, operation switch (panel screen), or dedicated remote control
Aufzeichnungsdatenformat
Kyowa standard file format KS2
Analysis using optional Data Analysis Software DAS-200A is possible.
Datensammlung
Online using PC, or offline direct reading from CF card to PC
TEDS
Only when using online control from a PC
Supported conditioner cards: CDV-40B (-F), DPM-42B (-F), DPM-42B-I (-F), CCA-40A (-F), CDV-44AS, CDA-44AS, CDA-45AS, CVM-41A
Betriebsluftfeuchtigkeit
20 –
90
% RHI
Vibrationsbeständigkeit
49.0 m/s2 (5 G), 5 to 55 Hz 1 cycle 1 min., each axis 15 cycles (Non-operating)
29.4 m/s2 (3 G), 5 to 55 Hz 1 cycle 1 min., each axis 15cycle (Operating)
Schlagfest
294.2
m/s2I
Standard Zubehör
DC power cable P-76 (1.8 m)
USB cable N-38 (1 m)
Ground wire P-72 (5 m)
CF card (1 GB) inserted in the slot
Fuses (8 A for 4-slot model, 5 A for 2-slot model)
Dummy panel
Installed on the free slots before shipment
EDX-200A-4H: 3 pcs
EDX-200A-2H: 1pc
EDX-200A-4T: None
EDX accessory bag
Dynamic data acquisition software DCS-100A (DVD)
Instruction manual (In English & Japanese, in the above DVD)
Optionales Zubehör
AC adapter UIA345-12 (For U.S.A.: UNI345-1238)
Fixing adapter
EDX dummy panel EDX1P-DUMMY
Remote control unit RCU-42A
Battery unit for instantaneous power failure EDB-41B
Monitor unit EMON-20A
Synchronous cables N-95, N-128