DT-A Displacement Transducers
(50 & 100 mm)
| Performance |
|
Non-linearity [%RO]
|
+/-0.5
|
|
Hysteresis [%RO]
|
+/-0.5
|
|
Repeatability [%RO]
|
0.3
max.
|
|
Rated output
|
1.5 mV/V (3000x10–6 strain) +/-20%
|
| |
| Environmental Capability |
|
Safe temperature range [°C]
|
0
–
60
(no dew condensation)
|
|
Compensated temperature range [°C]
|
0
–
50
(no dew condensation)
|
|
Temperature effect on zero balance [%RO/°C]
|
+/-0.05
|
|
Temperature effect on output [%RO]
|
+/-0.05
|
| |
| Electrical Characteristics |
|
Safe excitation voitage [V]
|
5
AC or DC
|
|
Recommened excitation voltage [V]
|
1
–
3
AC or DC
|
|
Input resistance
|
120 Ohm +/-3%
|
|
Output resistance
|
120 Ohm +/-3%
|
|
Cable
|
4-conductor (0.08 mm²) shielded chloroprene-coated cable of 4 mm diameter by 5 m long; terminated with an NDIS connector plug
|
| |
| Mechanical Properties |
|
Frequency response
|
DC to approx. 1.5 Hz
|
|
Measuring force
|
Approx. 4.4 N (reference value 450 gf)
|
| |
| Options |
|
Replacement Probes
|
X, XS, SH
|
|
Magnet Base
|
MB-B
|